Abnormal temperature and humidity display of constant temperature and humidity chamber.
The temperature and humidity controller of the constant temperature and humidity box shows that the temperature is abnormal. At this time, we must first check the super humidity protector in the control room to see if it is still at 150 ℃.
Secondly, check whether the pipeline in the laboratory has not been cleaned for too long, resulting in blocked water flow. Then check whether there is water in the upper water tank. If there is no water, check whether there is water in the lower water tank.
Finally, check whether the solid-state relay of humidity has a short circuit. If there is no problem with the heater, we can use the cross current voltage gear of the electric meter to detect.
You can also turn off the main power supply of the constant temperature and humidity box, open the drain valve under the heating barrel to a position parallel to the pipeline, discharge the water inside, clean the internal dirt, then pull out the silicone pipeline, scrub it in water, and install it back in the opposite direction after maintenance. After the installation, close the drain valve under the humidification bucket to a position perpendicular to the pipeline, and then turn on the main power supply behind the machine.
Frequent "machine tripping" occurs during the day, and the refrigeration effect is normal at night
How to solve the late overpressure:
Failure of damp heat system:
Features of programmable constant temperature and humidity test chamber: