Product Details:
|
Model Number: | LY-21000 | Volume: | 1000L |
---|---|---|---|
Usage: | Process Testing Machine, Precise New Benchtop Microclimate Test Chamber | Fluctuation: | ±0.5°C ±2.5%RH |
Temperature Range: | -40~150C | Humidity Range: | 20%~98% |
Test Hole: | D50mm | Material: | SUS 304# Stainless Steel |
High Light: | constant temperature and humidity chamber,environmental testing chamber |
PID Constant Climate Temperature Humidity Test Chamber With 1000 L Volume
Application
The Machine can simulate different environment condition. It's appropriate for testing material performance, such as resist heat, resist dry,resist humidity and resist cold. That can define performance of material.
Test Standard
GB/T2423/5170/10586, JIS C60068, ASTM D4714, CNS3625/12565/12566
Characteristics
• LED Touch Screen (Korea Temi)
• PID Control of Temperature and Humidity
• Both Temperature and Humidity are Programmable (can have 100pattern, each pattern has 999 segment)
• With Humidity Sensor
• With RS485/232 output to computer
• Window Software
• Remote fault notification(opntional)
• With Viewing Window
• With thermostats (prevent overheat)
• Cable Ports(50mm diameter)
• With Data Storage fuction by USB Flash Memory
• Protection (phase protection, overheat, overcurrent etc)
• Water Tank with level detector
• Adjustable Shelf
• Anti-condensation technology of the workroom .(Optional)
• User friendly three color LED indicator lamp ,easy to read working condition
Specification
Model | LY-280 | LY-2150 | LY-2225 | LY-2408 | LY-2800 | LY-21000 | |||
Inner Size W×H×D(cm) | 40×50×40 | 50×60×50 | 50×75×60 | 60×85×80 | 100×100×80 | 100×100×100 | |||
External Size W×H×D(cm) |
93×155×95 | 100×148×106 | 117×166×118 | 122×176×121 | 158×186×148 | 158×195×168 | |||
Volume(L) | 80 L | 150L | 225L | 408L | 800L | 1000L | |||
Temp&Humi Control Range |
Temp. Range: A: -20℃~150℃ Humi.range: RH20%-98% B: -40℃~150℃ C: -70℃~150℃ (Other temperature and humidity range can be customized) |
||||||||
Temp&Humi Fluctuation | ±0.5℃;±2.5%RH | ||||||||
Temp&Hum Departure | ±0.5℃-±2℃;±3%RH(>75%RH); ±5%RH(≤75%RH) | ||||||||
Controller Resolution | ±0.3℃;±2.5%RH | ||||||||
Controller Brand | Korean TEMI880 | ||||||||
Material | Inwall | SUS 304# steel plate | |||||||
Housing | SUS 304# steel plate surface fog processed | ||||||||
Heat Insulator | resistance to high temperature, high chlorine acid b vinegar foam insulation materials | ||||||||
Air Circulation Way | centrifugal fan-broadband type force air circulation | ||||||||
Refrigeration Way | cascade type multiple refrigeration | ||||||||
Refrigerater | fully enclosed piston compressor Taikang France | ||||||||
Refrigerants | R4O4Aor American dupont environmental protection cold media(R23+R404) | ||||||||
Condensing Way | Air-cooling | ||||||||
Heater | nickel chrome heating wire heater | ||||||||
Humidifier | semi-closed steam | ||||||||
Water Supply Way | automatic cycle water supply | ||||||||
Standard Accessories |
*50mm test hole on the left............................................................1PCS *PL inner lamp................................................................................1PCS *Adjustable clapboard.....................................................................2PCS *Humidity cloth...............................................................................1Bag *Fuses..............................................................................................3PCS *Power line.......................................................................................1PCS *Software CD....................................................................................1PCS *Operation manual.............................................................................1PCS |
||||||||
Safety Devices |
*Leakage current*Overload*Compressor overload*Over temperature*Ultra-wet/dry *Humidification temperature protection. |
||||||||
Power | 3phase 5 lines, 380V AC±10% 50/60Hz | ||||||||
Chamber structure Metal plate material: The inner and outside material is SUS mirror surface corrosion resistant plate, inner chamber two sides have sample Rail frame,it has the adjustable hang strip, heat preservation and heat insulation at least 10cm with hardness polyamine foaming. It has the high quality bakelite and glass fiber doorframe cutting off the heat conduction. Heat-proof quality is quite good,the door airtight seal strip is the two-way wide high-temperature resistance silicon foaming. Test space blowing-in circulate Mature blowing-in system use high temperature resistance multiple-blade wind wheel which is drived by the motor,the upside blade blow the wind in and the downside return the wind. When the temperature is -70-100℃,the error will be less 2℃ |
|||||||||
Controller system Use the Korea TEMI880 controller which is with low energy consumption (20W),the control system accuracy is ±0.3℃. High mark screen 5.7 inch LCD color screen with interactive parameter input method, resolution ratio is 800*480. Big memory capacity 120 groups / 1200 section all repeatedly:maximum is 999 (can repeated unlimited), part repeated: support maximum 99 PC-Link, MODBUS (ASCII, RTU). And has fact curve with temperature and humidity. Convenient and fast connect the computer Controller use R232 port,if you monitoring the history curve and need to playback the data,you can use editing operation. |
|||||||||
Heating and humidifying system Heater Spiral nichrome tubular electric heating element heater, High accuracy temperature fluctuation, and with long lifetime! Humidifier Use shallow slot method to humidify. Humidity measurment use wet and dry bulb method. Secondary protection Over-temperature protector:(protect the sample and the chamber) Humidifier empty burn preventer(protect the humidifier) |
Water level scale & water tank |
||||||||
Temp & Humi. Control Range(Specific test requirements customize) 1.Choice of wind circulation way: Forced air supply If you have requirements on wind supplying way, this kind of air supply is the best choice. 2. Choice of water supply system: RO (Reverse osmosis) Our equipment use pure water when doing humidity test, can automatically add water to make test more convenient. 3. Polymer film/wheel dehumidification system If you need to do low temperature and low humidity test for long time, then honeycomb chemistry dehumidifier system will be your best choice. It can make to 10℃ 10%RH or lower. 4. LN2/LCO2 fast cooling system We offer all kinds of imported professional LN2/LCO2 solenoid valve and related controlling system to meet your requirements of fast heating and cooling. 5. Equiped with glass door and operation hand hole on the door If you need to move or open the test sample during testing, to equip a glass door and operation hand hole to avoid effect to test condition when open the chamber door. 6. Recorder to get test data in time If need to record different points environmental and sample temperature, to equip a temperature recorder is a best choice. |
Contact Person: Victoria Li
Tel: +8613809275028
Fax: 86-769-28682030