Product Details:
|
Temp&Humi Fluct: | ±0.5℃;±2.5%RH | Temperature Contro: | A: -40℃~150℃,B: -60℃~150℃ ,C: -70℃~150℃ |
---|---|---|---|
Color: | White, Blue And No-painted | Volume: | 408L |
Inner Size(cm): | 60x85x80 | Humidifier: | Semi-closed Steam |
Highlight: | constant temperature and humidity chamber,environmental testing chamber |
800L White, Blue And No-Painted Laboratory Equipment Constant Temperature Humidity Climatic Test Chamber
Introduction:
Climate controlled cycling environmental temperature humidity test chamber for Lab is for testing material heat, cold, dry resistance, humidity resistance.Suitable for electronics, electrical appliances,Communications, instrumentation, vehicles, plastic products, metal, food,chemical,building materials, medical, aerospaceand other products.
Standards:
GB/T2423/5170/10586, JIS C60068, ASTM D4714, CNS3625/12565/12566 etc.
Characteristics:
1.Use the high precision microcomputer touch temperature and humidity controller with high stability of platinum resistance to temperature and wind speed in the circulatory system of temperature and humidity test.
2. Control temperature&humidity well-distributed, precisely and steady.
3.Completely independent system separate for testing in high temperature, low temperature and constant temperature&humidity environment.
Advantage of Temperature Humidity Control System Unit Test Equipment
● Small, portable, benchtop design
● Wide temperature ranges and volume selection.
● Programmable touch screen controller.
Specifications:
Model | LY-280 | LY-2150 | LY-2225 | LY-2408 | LY-2800 | |
Inner Size W×H×D(cm) |
40×50×40 | 50×60×50 | 60×85×80 | 60×85×80 | 100×100×80 | |
Outside Size W×H×D(cm) |
93×155×95 | 100×148×106 | 117×166×91 | 140×176×101 | 170×186×111 | |
Volume (V) | 80 L | 150L | 225L | 408L | 800L | |
Temp and Hum Range |
A:-20°C~150°C B: -40°C~150°C C: -60°C~150°C D: -70°C~150°C RH20%-98% |
|||||
Function | Fluctuation | ±0.5°C ±2.5%RH | ||||
Deviation | ±0.5°C-±2°C ±3%RH(>75%RH); ±5%RH(≤75%RH) | |||||
Controller Analytical Accuracy |
±0.3°C ±2.5%RH | |||||
Wind Cycling Way | Centrifugal fan-broadband type forced air circulation | |||||
Refrigeration Way | Single stage compression refrigeration | |||||
Refrigerator | French Tecumseh | |||||
Refrigerants | R4O4A USA DuPont environmental protection refrigerant(R23+R404) | |||||
Condensing Way | Air-cooled or water-cooled | |||||
Water Supplying Way | Automatic cycling water supply | |||||
Standard Accessories |
1 Observe window(double-deck cavity toughened glass), 1 50mm test hole in the left, 1 PL inner chamber lamp, 2 clapboards, 1 bag of wet and dry bulb, 3 fuses, 1 power line. |
|||||
Safety device |
Non- fuse-switch(compressor overload, refrigerant high low voltage, over-humidity and temperature protection, Protection switch,fuse stoppage warning system |
Contact Person: Victoria Li
Tel: +8613809275028
Fax: 86-769-28682030